China medical x ray machine manufacturer

printFavorites

ALL PRODUCTS

Home :: Products :: Medical X-ray Machine :: Conventional Radiography X-ray Unit :: 200mA Conventional Radiogrphy X-ray Unit YSX0108

200mA Conventional Radiogrphy X-ray Unit YSX0108

Product Type: Medical X-ray Machine

Place of origin: China

Model No: YSX0108

Price Terms: FOB, CIF

Payment Terms: L/C, D/P, T/T

Package:

Minimum Order:

Delivery Time:

Brand Name:

Products Informatin

200mA conventional radiogrphy x-ray unit YSX0108

Number: YSX0108
Main Characteristics of Medical X Ray Machine
● Single-table / Single-tube
● Rotary anode X-ray tube unit and tangential annular tube head
● Single-phase full wave rectification high-voltage generator
● Power voltage (V) radiograph kilovolt (KV). Step-less control electric mechanism
● Filament stabilizer for X-ray tube and space charge complementor
● Radiograph volume KV/mA and S. grading and interlock protection
● Adopt digital circuit timer. Grading according to R10 Priority coefficient. (for more accurately circuit of silicon-controlled rectifier of heavy power)
● Radiographic table can move in length and breadth.
● The radiographic table, column and vibrating grid are in a whole without ceiling and floor rails bottom
Main Technical Indexes of Medical X Ray Machine

Item Contents Index
Demand of Power Supply Capacity Two-phase 30KVA
Voltage Frequency 380V/220V ±10%  50HZ ±0.5HZ
Internal Resistance 0.75Ω (380V)  0.25Ω(220V)
Power Supply Adjustment Range 380±10% at 380V, 
220V±10% at 220V,  being adjustable continuously
Radiograph Voltage 50—125KV  being adjustable continuously
Current Small Focus 50mA 100mA
Big Focus 50mA 100mA 200mA
Time 0.04~6.35s  23grades together, according to R10 Coefficient
HT Generator Capacity 20KVA(Instantaneous)
Max DC Output Voltage 125KV
Max DC Output Currency 200mA
 X-ray Tube Model XD51. 20—40 / 125
Focus Small Focus 1×1 mm    Big Focus 2×2 mm
  Radiographic Table Length×Width×Height of Bed Surface 2000mm×665mm×700mm
Moving range of bed surface 640mm in Length, 200mm in breadth
Ray-filter of bed Travel in Length≥500mm
Grid Density N=28
Convergence Distance : f=100cm
 Grid Ratio :  R=8
Column of X-ray Tube Unit Moving Longitudinally along Radiographic Table 1000mm or 1800mm
Distance to ground as moving up and down (along column) 650~1800mm
Rotary around the center of cross arm ±180°
Rotary around the shaft line of X-ray Annular Tubes -10° ~ -60°~ +120°
Max Size of Cassette   356mm×432mm (14°×17°)

Note: We will not make another notice if the contour of product and technical parameter will be changed.